Total Nitrogen Analyzer Liquiline System: TN Analyzer for Liquid Samples

SUP Total Nitrogen Analyzer is a stable and accurate liquiline system analyzer that monitors the nitrogen content in various samples, like water, wastewater, soil, and other industrial materials. 

The TN Liquiline system analyzer is widely used in environmental monitoring, water treatment, agriculture, and industrial applications to assess nitrogen concentration, which is the key parameter for evaluating water quality, soil health, and compliance with environmental regulations.

Measuring range: (0~100 / 200 / 500 / 1000 / 2000) mg/L

Error:

  • 20% Range:±8%;
  • 50% Range:±5%;
  • 80% Range:±3%

Repeatability ≤3% 24h low concentration drift ±5 mg/L 24h high concentration drift

Characteristics

Low consumption of reagent: No more than 3mL of reagent in a single measurement

Separation of toxic and non-toxic wastewater: Super accurate and precise results are output based on the separation

Double optical path detection: Real-time calculation of absorbance, eliminating the interference during the measurement

Automatic turbidity compensation: Based on the turbidity of the sample, compensation is given automatically to avoid interference

Automatic range switching: Range adjusts automatically when measurement results exceed the range

Key data uploaded automatically: Key parameters are uploaded to assist in judgment based on the protocol requirements

Human-computer interaction: Linux system allows man-machine communication with a color touch screen

IPv4 network support: Compatible with IPv6 network and meets the communication requirements of TCP/IP protocol

USB and TF card support: Program upgrade, data export, and data storage upgrade allowed

  • Applications

① Monitoring the inlet and outlet of wastewater treatment plants for documentation and calculation of the measured liquid's cleaning ability 

② Providing the discharges of industrial wastewater treatment plants to determine discharge fees based on the polluter-pays principle

③ Evaluating surface water quality in medical units, river and canal water, laboratories, groundwater, ship wastewater, small watersheds, and coastal water

Total Nitrogen Analyzer Liquiline System: TN Analyzer for Liquid Samples

 

  • Measuring Principle

SUP Total Nitrogen Analyzer uses potassium persulfate under alkaline conditions to digest the sample at high temperature, and all the nitrogen contained in the sample will be oxidized into nitrate. It measures the absorbance of the digest sample at wavelengths of 220 nm and 275 nm, respectively, and it gets the absorbance A220 and A275. This accurate TN liquiline system analyzer calculates the end absorbance A according to the formula A = A220 - 2A275. On top of that, the instrument can calculate the concentration of total nitrogen in the sample through UV spectrophotometric conversion.

Total Nitrogen Analyzer Liquiline System: TN Analyzer for Liquid Samples

  • Specification

 Measuring range  (0~4 / 10 / 20 / 50 / 100 / 200) mg/L
 Error  20% Range:±8%; 50% Range:±5%; 80% Range:±3%
 Repeatability   ≤3%
 24h low concentration drift  ±5 mg/L
 24h high concentration drift  ≤3%
 Minimum Maintenance Period  ≥168 h/time
 Digital communication  RS232、RS485、RJ45
 Analog communication  (4~20)mA input; (4~20)mA output
 Power requirements  (220±22)VAC;(50±1)Hz
 Ambient temperature  (5~40)℃
 Size  300mm×420mm×240mm (W×H×D)
 Weight  <15 Kg

 

Total Nitrogen Analyzer Liquiline System: TN Analyzer for Liquid Samples