This high-precision measurement device features a 7.0-inch high-contrast LCD touchscreen and an intelligent operation system for easy management of users, calibration, methods, and data. It supports multiple reading modes—auto, timed, and continuous—and ensures accurate results with automatic or manual temperature compensation. The device stores up to 1,000 GLP-compliant data sets per parameter, provides data analysis for review and comparison, and offers USB or RS-232 connectivity. With IP54 waterproofing and flexible calibration options, it is designed for reliable and convenient laboratory or field use.
| Model |
EC600L |
| Parameters |
EC/Temp.(Resistivity/TDS/Sal.) |
| Conductivity |
Range |
0.000 μ S/cm to 3000 mS/cm |
| Resolution |
0.001 μ S/cm minimum, various with range selection |
| Accuracy |
± 0.5% FS |
| Reference Temperature |
5, 10, 15, 18, 20, 25 ºC |
| Resistivity |
Range |
5.00 Ω ·cm~100.00 M Ω ·cm |
| Resolution |
0.01 Ω ·cm minimum |
| Accuracy |
± 0.5% FS |
| TDS |
Range |
0.000 mg/L~1000 g/L |
| Resolution |
0.001ppm minimum, changed with range |
| Accuracy |
± 0.5%FS |
| Salinity |
Type |
NaCl% |
| Range |
(0.00~8.00)% |
| Resolution |
0.01% |
| Accuracy |
± 0.1% |
| Measurement |
Reading Mode |
Auto Read (Fast, Medium, Slow), Timed, Continuous |
| Reading Prompts |
Reading, Stable, Locked |
| Temp. Compensation |
ATC, MTC |
| Data management |
Data Storage |
1000 Groups |
| GLP Features |
Yes |
| Log Management |
Yes |
| Inputs |
Temp. Probe |
6-pin Mini-DIN |
| Conductivity Probe |
8-pin Mini-DIN |
| Outputs |
USB, RS 232 |
USB 2.0 flash memory device, printer, PC |
| Temperature Range |
-5 to 130 ºC |
| Power |
AC Adapter, 100-240V AC input, DC24V output |
| IP Rating |
IP54 |
| Dimensions |
280 × 280 × 130 mm |
| Weight |
2500g (5.51 lb) |
-
FEATURES
- Display: Color High contrast LCD touchscreen, 7.0 inches.
- System: Intelligent operation system provides features including user management, calibration management, method management, data management and etc.
- Reading: Multi-reading feature allows auto-read, timed-read and continuous-read.
- Compensation: Automatic/Manual temperature compensation ensures accurate results.
- Storage: Data Storage 1000 sets for each parameter (GLP-compliant).
- Analysis: Data analysis feature helps user review, compare and recalculate results.
- Communication: Support for USB or RS-232 communication.
- Reset: Reset feature automatically resumes all settings back to factory default options.
- Protection: IP54 waterproof.
- Calibration: 1-5 points calibration with Standard Recognition.
- Settings: Settable parameters, including cell constant, reference temperature (5/10/15/18/20/25°C), temperature compensation coefficient and TDS factor.
- Type: Temperature compensation type (none, linear, pure water).