SUP-PH5011 pH sensor

SUP-PH5011 pH sensor used in PH measurement is also called primary cell. The primary battery is a system whose function is to convert chemical energy into electrical energy. The voltage of this battery is called electromotive force (EMF). This electromotive force (EMF) consists of two half-cells.

Characteristics

Zero potential point: 7 ± 0.5 pH

Slope: > 95%

Installation size:  3/4NPT

Pressure: 0 ~ 4 Bar at 25 ℃

Temperature: 0 ~ 60℃

  • Specification

  Product 

  Plastic pH sensor

  Model

  SUP-PH5011

  Measurement range

  2 ~ 12 pH

  Zero potential point

  7 ± 0.5 pH

  Slope

  > 95%

  Internal impedance

  150-250 MΩ(25℃)

  Practical response time

  < 1 min

  Installation size

  Upper and Lower 3/4NPT Pipe Thread

  NTC

  NTC10K/Pt100/Pt1000

  Heat resistance

  0 ~ 60℃ for general cables

  Pressure resistance

  0 ~ 4 Bar

  Connection

  Low-noise cable

 

  • Introduction

SUP-5011 plastic ph sensor

  • Product advantages

Adopt international advanced solid dielectric and large area PTFE liquid junction, no clogging, easy maintenance.

Long distance reference diffusion path, extends electrode life greatly in harsh environments.

Using PPS / PC shell, Up and down 3/4NPT pipe thread, easy installation, no need sheath, saving installation costs.

Electrode is made of high quality low-noise cable, make signal output length greater than 40 meters or more, without interference.

No supplemental dielectric, a little maintenance.

High accuracy, fast response, good repeatability.

With silver ions Ag / AgCL reference electrode.

Proper operation to extend service life

Side or vertically installation to the reaction tank or pipe.

Related Products

Need more information?

Contact us for more details about our boilers,solutions and services